GE HealthCare AMX 4 Plus Forums

Updated  Feb 13 2024
2 Replies

Hello,

To start at the beginning ... 1st) Tube/collimator was tough to move up/down to column.  Replaced upper brake to column, brake wires shorted to housing when re-assembling.  Repaired wires, routed wires a different way, re-assembled ... Tube/collimator easy to move up/down.  Charged portable overnight, next day went to verify system working before giving to the techs .... rotor cuts out in 2 seconds & system restarts, tests passed, no errors, CB never trips off.   

2nd) Now AMX-4 + when you rotor from Hand switch, can hear relays click, 2 seconds later the displays go blank and startup test pass, display is ready again.   I can hear the rotor spinning in tube,  tube resistance is good. I verified capacitor values on right side. All fuses are good. Reseated Relays on circuit boards.  I have a replaced 1 board at time to the following .... Display Bd, CPU, Rotor Bd, FIL/Kvp Bd.  No errors appear after startup tests.  Can't do any cals because rotor doesn't last but 2 seconds.    Does anyone have any other idea's? 

Thanks

Jeremiah

Equipment: GE HealthCare - AMX 4 Plus

RE: GE HealthCare - AMX 4 Plus
Added  Jan 12 2024
0 Replies

I have facing problem when exposure button is pressed error papers 431 & 433 any help to resolvsolve this

RE: GE HealthCare - AMX 4 Plus
Updated  Jul 18 2023
1 Reply

Turned on AMX4 but test4 failed please help how to solve it 

Equipment: Amx4

RE: GE HealthCare - AMX 4 Plus
Updated  Apr 3 2023
1 Reply

How do I calibrate amx4plus

Equipment: Amx4plus

RE: GE HealthCare - AMX 4 Plus
Updated  Oct 26 2021
1 Reply

service manual needed

RE: GE HealthCare - AMX 4 Plus
Updated  Jul 19 2021
2 Replies

Hi, dear forum. I need the procedure to run a "LOOPING TEST" to an AMX-4 mobile system. Can anybody give me a hand with this issue?. I will be too thankful for your valuable help. Thanks before hand.

RE: GE HealthCare - AMX 4 Plus
Updated  Jun 23 2021
4 Replies

Hello, good morning we just get an amx the machine was working really fine, but after a while the batteries get dead, we tried to charge it back but it does not charge the display board dont show anything we tried to turn switch on to moe and it does get any charge. thank you, hope you help me

RE: GE HealthCare - AMX 4 Plus
0 Replies

Could you tell me what is the password required to enter to the Diagnostic items, for example Loop TEST, IN THE AMX-4 SERIES ?

RE: GE HealthCare - AMX 4 Plus
Updated  Jun 17 2021
2 Replies

Hello, GE AMX 4 Plus. During system turns on it shows TEST 04 Failed, after finally loading shows DRIVE ONLY, X RAY DISABLED. In the service there is CHECKSUM ERROR 11, during FILAMENT CURRENT TABLE CALLIBRATION, shows ERROR 432. Please take an advice, is this issue from HV TANK, TUBE or FIL/KVP CONTROL BD, Who had solved issue like this.

RE: GE HealthCare - AMX 4 Plus
Updated  May 12 2021
1 Reply

Trying to do an exposure, the display show Error 439 and Generator Fault messages

RE: GE HealthCare - AMX 4 Plus

Start a Thread

Equipment Name: GE HealthCare - AMX 4 Plus
SIGN IN FIRST
This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies. Please review our Privacy Policy for more details.
I Agree