Hello,
To start at the beginning ... 1st) Tube/collimator was tough to move up/down to column. Replaced upper brake to column, brake wires shorted to housing when re-assembling. Repaired wires, routed wires a different way, re-assembled ... Tube/collimator easy to move up/down. Charged portable overnight, next day went to verify system working before giving to the techs .... rotor cuts out in 2 seconds & system restarts, tests passed, no errors, CB never trips off.
2nd) Now AMX-4 + when you rotor from Hand switch, can hear relays click, 2 seconds later the displays go blank and startup test pass, display is ready again. I can hear the rotor spinning in tube, tube resistance is good. I verified capacitor values on right side. All fuses are good. Reseated Relays on circuit boards. I have a replaced 1 board at time to the following .... Display Bd, CPU, Rotor Bd, FIL/Kvp Bd. No errors appear after startup tests. Can't do any cals because rotor doesn't last but 2 seconds. Does anyone have any other idea's?
Thanks
Jeremiah
Equipment: GE HealthCare - AMX 4 Plus
RE: GE HealthCare - AMX 4 PlusI have facing problem when exposure button is pressed error papers 431 & 433 any help to resolvsolve this
RE: GE HealthCare - AMX 4 PlusTurned on AMX4 but test4 failed please help how to solve it
Equipment: Amx4
RE: GE HealthCare - AMX 4 Plus