STERIS - AMSCO 3080 R
by STERIS

STERIS AMSCO 3080 R Forums

Updated  Mar 30 2018
6 Replies

Would anyone have an idea as to how long the batteries should last for this table?

RE: STERIS - AMSCO 3080 R
Updated  Mar 4 2016
1 Reply

WOULD ANYONE  HAVE  A P/M PROCEDURE  FOR  THESE BEDS? 

 

RE: STERIS - AMSCO 3080 R
Updated  Mar 10 2016
7 Replies

I have a 3080 RL that has a lot of play in the top. Meaning the entire top section of the table "wiggles" very easily and feels rather loose. I have spoken with Steris about this before and they told me its just a result of age. This makes sense however Im curious if anyone else has figured out how to solve this issue.

RE: STERIS - AMSCO 3080 R
Updated  Sep 23 2016
8 Replies

Having an issue with the table not running off the batteries. It will run plugged into the wall. I disconnected the batteries and measured the voltage on each at 13v fully charged. (12 volt battery) To me it seems like a control board issue. Has anyone else experienced this? Anyone have a service manual? I have a 3085 manual which is close but I am not sure if there are any big differences. 

 

 

RE: STERIS - AMSCO 3080 R
Updated  Jun 17 2015
3 Replies

I have a 3080 RL/SP (BE587283 actual model) with multiple problems: blown fuse in fuse drawer, shorted full wave bridge, flaky circuit breakers, bad motor batteries, auxiliary switches activate relays but not solenoids (even with good motor batteries jumped in). If someone could respond in this forum with suggestions or an e-mail/ phone number I could reply to so that we may discuss the items noted, it would be greatly appreciated.

RE: STERIS - AMSCO 3080 R
Updated  Apr 15 2015
2 Replies

Hi, we are looking for P764328571 rplacment kit

for our broken RC.

 

If someone know where to buy for reasonable price please contact me.

 

Thanks

RE: STERIS - AMSCO 3080 R
Updated  Jan 30 2014
2 Replies

Has any body had a problem with a Steris surg. table activating by itself. No buttons being pushed, that is what I am being told.

RE: STERIS - AMSCO 3080 R
Updated  Jul 23 2013
6 Replies

Have Steris 3080 that is not charging batteries and no AC power indication on hand-control. Unit fuctions normally when plugged into to ac & using the side toggle switches. The power supply has been replaced.

RE: STERIS - AMSCO 3080 R
Updated  Jul 16 2012
1 Reply

Need a Steris 3080R OR table service manual

RE: STERIS - AMSCO 3080 R

Start a Thread

Equipment Name: STERIS - AMSCO 3080 R
SIGN IN FIRST
This site uses cookies. By continuing to browse the site you are agreeing to our use of cookies. Please review our Privacy Policy for more details.