Hi all,
I have the panda resus hooked up to a liner and canister. When I turn the suction on, I'm expecting a vacuum of 50 mmHg but only measure ~20 mmHg. I also try increasing the expected vacuum but it still only measures ~20 mmHg. The vacuum does increase when the liner is manually pushed down when the suction is on, but manual intervention shouldn't be necessary.
The resus passed all checkout tests, I tried calibrating it anyways, and it passed all checkout tests again.
I checked for cracks or leaks in the liner and canister and didn’t notice any. Tried replacing the liner and canister with another set and that didn't resolve my problem. I also tried both the original set and the replacement set on another vacuum and they both worked as expected without manual intervention.
Any recommendations for what to try next? Tests, replacement parts, or something else?
RE: GE HealthCare - Panda iRESHello all,
The screen on my Panda will not light up.
I can see the screen when I shine a flashlight on it.
Is it just the backlight or is there something else involved?
RE: GE HealthCare - Panda iRESEquipment: GE HealthCare - Panda iRES
RE: GE HealthCare - Panda iRESI have a panda that the procedure lights intermittently turn off.
Equipment: GE Healthcare - Panda iRES
RE: GE HealthCare - Panda iRESI have a Panda warmer that the bed will not go up or down. I have tried the pedals on both sides, still no response. However, I when I press the pedals, I do hear a relay clicking. Please help.
Equipment: GE Healthcare - Panda iRES
RE: GE HealthCare - Panda iRESEquipment: GE Healthcare - Panda iRES
RE: GE HealthCare - Panda iRESWhen I turn on the machine in pre-warming everything is ok bur in like 10 min working when the system is already warm the observations lights suddenly shut off and I can not turn in on again until the unit is cool again.
RE: GE HealthCare - Panda iRES